Model 9130 Triple Output Programmable DC Power Supply INSTRUCTION MANUAL MODEL 9130

Seite 2 - Safety information

10 14) Verify channels 2 & 3 according to steps 1) – 8) 5.If the Power Supply Does Not Turn On Use the following steps to help resolve problem

Seite 3 - TABLE OF CONTENTS

11 arrow on the handle and the arrow on the plastic ears are in opposite directions. Now you can pull the handle outwards. After removing the handle,

Seite 4

12 3. Specifications 3.1 Specifications Parameter 9130 Output ratings Voltage 0~30V×2, 0~5V×1 Current 0~3A×2, 0~3A×1 LVP* 0~31V×2, 0~6V×1 Load Regul

Seite 5 - 1. Introduction

13 8.45in(W) x 3.8in(H) x 13.9in(D) *) LVP: Limit Voltage Protection. Limits the voltage than can be set either via the front panel or remote con

Seite 6 - Front panel

14 4. Front-panel Operation So far we have covered the quick start chapter which briefly introduced the front panel operation and how to check basic

Seite 7 - 2.2 Preliminary Checkout

15 4.2 Panel Description 41 27536Esc0●98RecallMenuEnterV-setV-setI-setI-setSaveOn/Off 0~9

Seite 8

16 4.4 Menu Descriptions Press Menu to indicate operation mode. View the menu on the VFD, and use the keys or the knob to scroll through the com

Seite 9 - Lost Factory Calibration

17 Power SN display the serial number Soft Version Display the firmware version Cal Information Error Information Exit Exit Menu 4.5 Pan

Seite 10 -

18 Solution 2: Press V-setV-set, then use the arrow keys to move the cursor to the desired position and adjust the value with the knob. (Make sure

Seite 11

19 Press Recall followed by a number 1 – 50 to recall operating state assigned to this location. You can also use the SCPI command:*SAV、*RCL to sav

Seite 12 - 3. Specifications

2 Safety information Please review the following safety precautions before operating our equipment. General information The following safety precaut

Seite 13

20 Baud Rate Set This parameter configures the baud rate for serial communication. Possible values are 4800,9600,19200 or 38400. When operating

Seite 14 - I-setI-set

21 Out Series Set This function configures the instrument for series operation of 2 channels. “None” means that each channel operates independently

Seite 15

22 Connection diagram example: Channels 1 and 2 connected in parallel Note: When configuring for series or parallel operation, the user must w

Seite 16 - 4.4 Menu Descriptions

23  Power Information The following menus provide information about the specific power supply unit. Power Model: the model of power supply For exam

Seite 17 - Timer Operation

24 5. Remote Operation The DB9 interface connector on the rear panel of the power supply can be connected to a RS-232 or USB interf

Seite 18 - Set Current

25 5.2 Communication between Power Supply and PC Before putting the instrument into remote operation, make sure that the baud rate, parity bit and c

Seite 19 -  Power Config

26 5.3 SCPI Command Overview 5.3.1 Common IEEE488.2 commands *CLS *ESE *ESE? *ESR? *IDN? *OPC *OPC? *PSC *PSC? *RST *SRE *SRE

Seite 20 -  System Set

27 CONDition? INSTrument [:SELect] {FIRst|SECOnd|THIrd} [:SELect]? NSELect {1|2|3} NSELect? OUTPut [:STATe] {0|

Seite 21

28 :RWLock DISPlay: [:WINDow][:STATe] {OFF|ON} [:WINDow][:STATe]? MEASure[:SCALer] :CURRent[:DC]? [:VOLTage][:DC]?

Seite 22 - Out Time Set

29 APPly:CURRent[:LEVel][:IMMediate][:AMPLitude]? This command is used to read out the 3 channels’ current setting value at the same time. Parameter

Seite 23 -  Power Information

3 TABLE OF CONTENTS 1. INTRODUCTION ...

Seite 24 - 5. Remote Operation

30 MEASure[:SCALar]:POWer:ALL[:DC]? This command is used to read out the measured output power value of the 3 channels simultaneously. Parameter type

Seite 25

31 The Structure of the condition register is as followed: condition event enable condition

Seite 26 - 5.3 SCPI Command Overview

32 5.4.2 SCPI Command Description Common IEEE488.2 Commands *CLS This command clears the following registers:  Standard event status register

Seite 27 - CALibration

33 VX. XX software version number For example: BK PRECISION, 9130, 000000000000111101,V1.66 *OPC When all commands sent t

Seite 28

34 parameters include constant current, constant voltage, maximum voltage value and step voltage values. Command syntax: *SAV Parameter: 0~

Seite 29

35 Example: SYST:RWL SYSTem:ADDRess This command reads the communication address of the power supply. Command syntax: SYSTem:AD

Seite 30

36 Parameter: 0~255 Reset value: consult PSC command Example: STAT:OPER:ENAB 110 STATus:OPERation:ENABle? This command querie

Seite 31

37 STATus:OPERation:INSTrument:CONDition? This command queries parameters from operation condition sub register. After this command is executed, th

Seite 32

38 Parameter: 0~999999S (if you set 0S,the output-timer will be disabled.) Reset value: It is concerned about the parameter you set in the

Seite 33

39 [SOURce:]VOLTage:PROTection[:LEVel][:IMMediate][:AMPLitude] This command sets the max. voltage for the current channel. Command syntax: [SO

Seite 34

4 6.5 Test sequence generation ... 44 6.5.1

Seite 35 - STATus:QUEStionable:ENABle

40 6. PV9130 Software Guide 6.1 Introduction PV9130 is a software application for the 9130 Triple Output Programmable DC Power Supply. To use this

Seite 36

41 1234567891011 1 Configure the software operating parameters and serial communication parameters. 2 Voltage chart 3 Voltage and current d

Seite 37

42 6.3.1 Save and Open Save the program, Quick Set, voltage sweep and GO/NG settings to a PAR file. Open the program, Quick Set, voltage sweep a

Seite 38

43 6.3.3 Status bar The status bar will give you the communication information. When communication is successful, the status bar will display the fol

Seite 39

44 6.4.2 Setting Voltage and Current Before setting voltage/current values, you need to select a channel. If you want to set voltage and current for

Seite 40 - 6. PV9130 Software Guide

45 Voltage Sweep Before this operation, please select the channel you need: . Enter parameters StartValue, StopValue, voltageStep and Delay. Press

Seite 41

46 [left intentionally blank]

Seite 42

47 Service Information Warranty Service: Please return the product in the original packaging with proof of purchase to the address be

Seite 43 - 6.4 Operation

48 22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2014 B&K Precision Corp. V060214 Print

Seite 44

5 1. Introduction Description The 9130 is a fully programmable triple Output DC Power Supply delivering 0-30V/0-3A on 2 outputs and 0-5V/0-3A on 1 o

Seite 45 - 6.5.2 Go/NG

6 2. Quick Start This chapter will help you getting acquainted with the front panel operation of this power supply. 2.1 Front & Rear Panel Front

Seite 46

7 Rear Panel 1 2 3

Seite 47

8 2.Connect the power cord and turn on the power supply When you turn on the power supply, the front-panel display will light up briefly while the po

Seite 48

9 Note: “X” denotes the channel for which calibration data was lost. If the calibration data in the EEPROM is corrupted or the factory calibration v