Examples: :VOLT:LOW?
Response: NRf
CONSIDERATIONS:
1) The high level must be greater than the low level.
2) The difference between the levels must conform to 0.5V difference 10V
3) The low level may not be less than the low limit.
4.12.1.4 Predefined High Voltage Level
This command is used to set the predefined high level of the pulse. The pulse will be set when the
predefined USER levels are invoked to this high level.
COMMAND TYPE: Setting or Query
SETTING
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:PHIGH<ws>
<predef high level>[units]
Arguments:
Type: NRf
Units: MV, V (default)
Range: -9.5V to +10V
Rounding: To 10mV
Example: :VOLT:PHIGH 4V
QUERY
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:PHIGH?
Example: :VOLT:PHIGH?
Response: NRf
4.12.1.5 Predefined Low Voltage Level
This command is used to set the predefined low level of the pulse. The pulse will be set when the
predefined USER levels are invoked to this low level.
COMMAND TYPE: Setting or Query
SETTING
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:PLOW<ws>
<predef low level>[units]
Arguments:
Type: NRf
Units: MV, V (default)
Range: -10V to +9.5V
Rounding: To 10mV
Examples: :VOLT:PLOW 4V
QUERY
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:PLOW?
Examples: :VOLT:PLOW?
Response: NRf