Inhaltsverzeichnis

Seite 1 - Generator

Model: 4045B Arbitrary Function Generator USER MANUAL

Seite 2

10 3.6.4 SWEEP Key ... 34 3.6.5 MODULATION Key ... 36 3.7 ON Key ...

Seite 3 - Safety Summary

100 :STARt <numeric value> :LENGth <numeric value> :SAVe Point Rate ARBitrary:PRATe <point rate> This command i

Seite 4 - WARNINGS AND CAUTIONS

101 Address :ARBitrary:ADDRess <address> This command sets the current address of the waveform. It is used to determine to where arbi

Seite 5

102 Type: Numeric. Definite form arbitrary block. Indefinite form arbitrary block Numeric Range: -2047 to

Seite 6 - Compliance Statements

103 Considerations: 1) The value of the data at the start and end points must first be set by the user, using the :ARB:DATA command.

Seite 7 - Safety Symbols

104 Range SIN: 16 to 1000; divisible by 4 SQU: 2 to 1000; divisi

Seite 8 - CE Declaration of Conformity

105 TRI 0 NOIS 0 Start Address :ARBitrary:STARt <start address> This command sets the start address of the wave

Seite 9 - 3 OPERATING INSTRUCTIONS

106 This command sets the length of the waveform being run. Arguments Type: Numeric

Seite 10

107 SERVICE INFORMATION Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return

Seite 11

108 LIMITED THREE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will

Seite 12 - 1 Introduction

109 Index Amplitude, 13, 28, 50 cursor, 49 duty cycle, 28 Errors, 51 factory default, 50 frequency, 28 Impedance Matching, 26 LCD, 24 Linear, 35 O

Seite 13 - 1.3 Specifications

11 4.3.1 The Input Buffer ... 68 4.3.2 The Output Queue ... 68 4.3.3 Response Messages ...

Seite 14

22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2012 B&K Precision Corp.

Seite 15

12 1 Introduction 1.1 Introduction This manual contains information required to operate the B&K Precision model 4045B Arbitrary Function G

Seite 16

13 1.3 Specifications Model 4045B Frequency Characteristics Sine 0.01 Hz – 20 MHz Square 0.01 Hz – 20 MHz Triangle 0.01 Hz – 2 MHz Accuracy 0

Seite 17

14 Square Rise/Fall Time < 20 ns (10% to 90% at full amplitude into 50 Ω) Variable Duty Cycle/Symmetry Square: 20% to 80 %, up to 2 MHz Triangle:

Seite 18 - 2 Installation

15 – 10 s) Modulation Characteristics Amplitude Modulation Internal 0.1 Hz – 20 kHz sine, square, or triangle waveform Ext

Seite 19 - 2.4 Power Requirements

16 Accuracy ± 0.02 % ± 2 digits Sensitivity 25 mVrms typical General Memory Storage Store up to 20 instrument settings Arbitrary memory 1,000 po

Seite 20 - 2.6 Grounding Requirements

17 To ensure the most current version of this manual, please download the latest version here: http://www.bkprecision.com/search/4045B For cu

Seite 21 - 2.7 Signal Connections

18 2 Installation 2.1 Introduction This section contains installation information, power requirements, initial inspection and signal connectio

Seite 22 - 3 Operating Instructions

19 Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision. 2.3 Instrum

Seite 24 - 3.4 Connectors

20 2.5 Fuse Replacement There is a 1A, 250V rated slow blow fuse at the AC input. Should the fuse ever get blown, follow the steps below to repla

Seite 25 - 3.5 Output Connections

21 WARNING TO AVOID PERSONAL INJURY DUE TO SHOCK, THE THIRD WIRE EARTH GROUND MUST BE CONTINUOUS TO THE POWER OUTLET. BEFORE

Seite 26

22 3 Operating Instructions 3.1 General Description This section describes the displays, controls and connectors of the function generator. All

Seite 27 - 3.6 MENU Keys

23 1 Power Button Power ON/OFF unit 2 LCD Display Displays all instrument data and settings 3 Function Keys F1 – F4 function keys to select

Seite 28

24 3.2 Display Window The function generator has a color LCD display that can display up to 400 x 240 dots. When powering on the unit, sine wavef

Seite 29

25 delivered from the function generator. Output Connector Use this connector to transfer the main output signal from the function generator.

Seite 30 - Mode Menu

26 Excessive distortion or aberrations caused by improper termination are less noticeable at lower frequencies, especially with sine and triangle wa

Seite 31

27 3.6 MENU Keys These keys select the main menus for displaying or changing a parameter, function, or mode. 3.6.1 WAVEFORM Keys These

Seite 32 - 3.6.3 UTILITY Key

28 F1: Frequency – Selects and displays the frequency. Change the frequency setting using the cursor keys, rotary knob, or numerical keys. F2: Amp

Seite 33

29 Square Menu 3.6.2 MODE Key Selects the output mode: CONT (Continuous), TRIG (Triggered), GATE (Gated), and BRST (Burst). To select

Seite 34

3 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of opera

Seite 35

30 Mode Menu F1: Continuous – Selects continuous output. F2: Triggered – Triggers one output cycle of the selected waveform for each trigger even

Seite 36

31 Trigger Menu F1: Manual – Selects manual as the trigger source. To trigger the waveform generator, press this MANUAL trigger button again.

Seite 37

32 Burst Menu 3.6.3 UTILITY Key Utility Menu

Seite 38 - FM Menu

33 F1: Recall – Recalls a previously stored front-panel setup from the selected buffer. Change the buffer number by using the rotary input knob

Seite 39

34 Counter Screen Press F1 - Off to turn off the counter. 3.6.4 SWEEP Key Selects the Sweep Mode and allows entering of sweep parameters: Swee

Seite 40

35 Sweep Menu F1: Off – Disables the sweep function. F2: Linear – Selects the Linear sweep shape. F3: Logarithmic – Selects the Logarithm

Seite 41

36 Set Sweep Menu 3.6.5 MODULATION Key Selects the AM or FM Modulation mode. To select the output mode, press MODUL key, then press the fun

Seite 42

37 Modulation Menu Press F2 to select AM menu: AM Menu

Seite 43

38 F1: % - Defines the modulation depth (from 0 to 100%) F2: Frequency - Selects the modulation frequency, from 0.1 Hz to 20.00 kHz

Seite 44

39 F1: Deviation - Defines the FM deviation frequency. F2: Frequency - Selects the modulation frequency, from 0.1 Hz to 20.00 kHz. F3: Shape

Seite 45

4 adjustments must be made by qualified maintenance personnel. Disconnect the power cord before removing the instrument covers and replacing compo

Seite 46

40 F1: Frequency Rate - (Frequency) Selects and displays the frequency. Change the frequency setting using the cursor keys, rotary knob or numerica

Seite 47

41 waveform generator will display a “Setting Conflict” error message. F4: Arb - Selects the Arbitrary waveform editing menu: Arbitrary Ed

Seite 48 - 3.7 ON Key

42 waveform, a message to save the Arbitrary wave will be displayed. Select YES or NO to save the new waveform. 3.6.7 Arbitrary EDIT Menu Ent

Seite 49 - 3.10 Power-On Settings

43 Edit Menu F1: Point - This menu allows point-by-point waveform editing. When selected, the following menu is displayed: Point Menu

Seite 50 - 3.11 Memory

44 F1: Adrs - Select the current address in the arbitrary waveform memory. F2: Data - Selects the data point value at the current addr

Seite 51 - 3.12 Displaying Errors

45 F1: From - Selects the starting point address. F2: To - Selects the ending point address. F3: Exec - Displays the Confirmation me

Seite 52 - 3.13 Quick Start

46 Predefine Menu F1: Type - Use the rotary dial to select the waveform Sine, Triangle, Square or Noise. If Noise

Seite 53

47 waveform (number of points for a full wave). Different waveforms have different limitations on the length, as shown in Table 3-1. Table

Seite 54

48 function menu options, ADD and NEW are available. Select ADD to add noise to an existing waveform, or NEW to create a new noise waveform. F4:Sh

Seite 55

49 3.8 Cursor Keys Use these keys to move the cursor (when visible) either left or right. They are used in conjunction with the rotary input

Seite 56

5 WARNING: To avoid electrical shock hazard, disconnect power cord before removing covers. Refer servicing to qualified personnel. CAUTION: Befo

Seite 57

50 assumes power-on default settings. Table 1 lists the factory default settings or selected after RECALL 0. Table 1 – Power-on Default Settings K

Seite 58 - (Examples)

51 3.12 Displaying Errors The waveform generator displays error messages when front-panel settings are either invalid or may produce unexpect

Seite 59

52 3.13 Quick Start This section explains how to generate various waveforms and modify the output waveform. * Generating a waveform output

Seite 60 - 4 Programming

53 3.13.3 Using Voltage Offset Through the offset parameter you can add a positive or negative DC level to the output waveform. To set vo

Seite 61

54 The function generator will overwrite and store settings into a buffer that had settings previously stored inside without a warning. Recalling S

Seite 62

55 3.13.5 Creating an Arbitrary Waveform You can create an arbitrary waveform using the following methods: * Enter individual data points

Seite 63

56 memory, follow these steps: 1. Press ARB main key to display the selection menu. 2. Press F4 :ARB to display the arbitrary menu. 3.

Seite 64 - For Windows XP:

57 execution point rate is the execution time between each point in the waveform. The total time taken to run one period of the waveform is giv

Seite 65

58 3.13.8 Setting the Amplitude The following equation represents the relative output amplitude voltage relationship between the front-panel ampl

Seite 66

59 2. Press the ARB key and select the F4:ARB function. 3. Press F1:START to set the address. Valid entries range from 1 to 999. 4. Press F2:LE

Seite 67 - 4.2 Device States

6 Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with

Seite 68

4 Programming 4.1 Overview This section provides detailed information on programming the generator via the USB (virtual COM) interface. 4.

Seite 69 - 4.6 Command Syntax

61 4. Click the “Close” button to stop the automatic search of the driver from Windows Update. 5. Now, go to Device Manager on the computer (Ope

Seite 70

62 7. In the following window, select “Browse my computer for driver software”, and following this, select “Let me pick from a list of device driv

Seite 71

63 10. Now click “OK”. It will show in the window under Model “AT91 USB to Serial Converter”. Click “Next” and you will get the following note:

Seite 72

64 For Windows XP: 1. The USB driver is included in the CD that comes with the instrument. You can also go to www.bkprecision.com and browse this

Seite 73

65 downloaded from the website labeled “atm6124_cdc.inf”. Select it and click “OK”. 9. The following screen will appear. 10. Click “Next”, and a

Seite 74 - 4.7 Status Reporting

66 11. Click on “Continue Anyway” and the driver will now install. 12. The driver will now install. Once finished, under “Device Manager” (This ca

Seite 75

67 BAUDRATE: 115200 PARITY: NONE DATA BITS: 8 STOP BIT: 1 FLOW CONTROL: NONE 4.2 Device States The device may be in one of the two possibl

Seite 76

68 4.3.1 The Input Buffer The device has a 128-byte long cyclic input buffer. Decoding of remote messages is begun as soon as the input buffer is

Seite 77

69 4.4 Instrument Identification The *IDN? common query is used to read the instrument's identification string. The string returned i

Seite 78

7 Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Chassis

Seite 79

70 The Program Message Unit can be divided into three sections as follows: a) Program Header The Program Header represents the operatio

Seite 80 - 4.8 COMMON COMMANDS

71 The Program Header Separator is used to separate the program header from the program data. It consists of one or more white space characters, de

Seite 81

72 iii) NRf This is a decimal numeric data type, where NR1 indicates an integer number, NR2 indicates a fixed point real number, an

Seite 82 - COMMANDS

73 For example, to set the frequency to its maximum value we can send the command FREQ MAX Some Program Message Units either require,

Seite 83

74 will return the maximum value to which the frequency may currently be set. Not all Program Message units have query forms (for example, SAV)

Seite 84

75 The first error in the queue is returned, and the queue is advanced. 4.7.2 Error Codes The negative error codes are defined by SCPI. Positiv

Seite 85

76 More parameters than allowed were received -109 Missing parameter Fewer parameters than necessary were received -110 Command he

Seite 86

77 -178 Expression data not allowed Execution Errors An execution error indicates that the device could not execute a syntactica

Seite 87

78 The parameter value must be selected from a finite list of possibilities. Device-Specific Errors An error specific to the device o

Seite 88

79 -410 Query INTERRUPTED. Data were sent before the entire response of a previous query was read. -420 Query UNTERMINATED. An at

Seite 89

8 CE Declaration of Conformity The model 4045B meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatib

Seite 90

80 4.8 COMMON COMMANDS 4.8.1 System Data Commands *IDN? - Identification query The identification query enables unique identification of the

Seite 91

81 4.8.3 Device Trigger Commands a) *TRG - Trigger command This command is analogous to the IEEE 488.1 Group Execute Trigger interface mes

Seite 92

82 *SAV - Save instrument state This command is used to store the current instrument state in the specified memory location. Arguments

Seite 93

83 specifications. The '<ws>' is used to denote a white space character. Note: When controlling the instrument remotely, do not i

Seite 94

84 SWRAte <numeric value> SWSTArt <numeric value> SWSTOp <numeric value> MODE CONT/ TRIG / GATE / BRST TRIG

Seite 95

85 FREQ 5E3 FREQ MAXIMUM FREQ MIN

Seite 96

86 Rounding: to 4 digits Command Type: Setting or Query Setting S

Seite 97

87 AMPL 2.5V AMPL MAX Query

Seite 98

88 OFFSet<ws>MINimum|MAXimum Examples: OFFS 2.5 OFFS 2.5V

Seite 99

89 DCYCle <duty cycle value> This command is used to set the duty-cycle of the square wave or the symmetry of triangular wave. The value

Seite 100 - Point Rate

9 1 INTRODUCTION ... 12 1.1 Introduction ... 12 1.2 Description ...

Seite 101 - Address

90 MODULation This command activates or deactivates modulation: Command Type: Setting or Query Setting

Seite 102 - Line Draw

91 Examples: DEPTh 50 Query Syntax: DEPTh?[<ws>MINimum|MAXimum]

Seite 103 - Predefined waveforms

92 MODSHape This command selects the modulating waveform shape Arguments Type: Character

Seite 104

93 DEV 5E3 DEV MAXIMUM DEV MIN

Seite 105 - Wavelength

94 SWRAte This command sets the time for one complete sweep: Arguments Type: Numeric

Seite 106

95 SWSTArt 5E3 SWSTArt MAXIMUM

Seite 107 - SERVICE INFORMATION

96 MODE <trigger mode> This command is used to set the trigger mode. Arguments Type:

Seite 108 - LIMITED THREE-YEAR WARRANTY

97 Setting Syntax: TRIGger<ws><INT|EXT> Examples: TRIG EXT

Seite 109

98 TRAte <trigger rate> Sets the rate of the internal trigger. Arguments Type: Numeri

Seite 110

99 Query Syntax: ERRor? Response: <Error number>, "<error description>" 4.9.2 Ar