OPERATING INSTRUCTIONS
MODEL 2703C
DIGITAL MULTIMETER
VV
mA
A
BATT 1.5V
COM V
10A
2703C
SAFETY INFORMATION
The following safety information must be observed to ensure maximum personal
safety during the operation at this meter:
Use themeter only as specified in this manual or the protection provided by the
meter might be impaired.
Test the meter on a known voltage before using it to determine if hazardous
voltage is present.
Do not use the meter if the meter or test leads look damaged, or if you suspect
that the meter is not operating properly.
Never ground yourself when taking electrical measurements. Do not touch
exposed metal pipes, outlets, fixtures, etc., which might be at ground potential.
Keep your body isolated from ground by using dry clothing, rubber shoes, rubber
mats, or any approved insulating material.
Turn off power to the circuit under test before cutting, unsoldering, or breaking
the circuit. Small amounts of current can be dangerous.
Use caution when working above 60V dc or 30V ac rms. Such voltages pose a
shock hazard.
When Using the probes, keep your fingers behind the finger guards on the
probes.
Measuring voltage which exceeds the limits of the multimeter may damage the
meter and expose the operator to a shock hazard.Always recognize the meter
voltage limits as stated on the front of the meter.
SPECIFICATIONS
Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999.
Polarity:Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed.
Zero:Automatic.
Lowbattery indication: The
" "
is displayed when the battery voltage drops
belowthe operating level.
Measurement rate: 2.5 times per second, nominal.
Auto power off:Approx. 25 minutes.
Operating environment: 0to 50at < 70% relative humidity.
Storage temperature: -20to 60, 0 to 80% relative humidity.
Accuracy: Stated accuracy at 23±5, < 75% relative humidity.
Temperature Coefficient: 0.1 x (specified accuracy) per . (to 18, 28 to
50).
Altitude: 6561.7 feet (2000m).
Power: Single standard 9-volt battery, NEDA1604, JIS 006P, IEC 6F22.
Battery life: 200 hours typical with carbon-zinc.
Dimensions: 165mm (H) x78mm (W)x42.5mm (D).
Weight:Approx. 10.0 oz.(285g) including holster.
Accessories: One set test leads, one spare fuse, 9V battery (installed), and
Operating Instructions.
DC VOLTS
Ranges: 2V, 20V, 200V, 1000V
Resolution: 1mV
Accuracy: ±(1.2% rdg + 1 dgt)
Input impedance: 10MΩ
Overload protection: 1000VDC or750VAC rms
AC VOLTS (50Hz - 500Hz)
Ranges:200V, 750V
Resolution: 0.1V
Accuracy: ±(2.0% rdg + 5dgts)on 200V range
±( 2.0% rdg + 10dgts) on 750V range
Input impedance: 10MΩ
Overload protection: 1000VDC or750VAC rms
DC CURRENT
Ranges: 2mA, 20mA, 200mA, 10A
Resolution: 1uA
Accuracy: ±(1.5% rdg + 1 dgts)on 2mAto 200mAranges
±(3.0% rdg + 3 dgts) on 10Arange
Input protection: 0.5A/500V fast blow ceramic fuse on mAinput
10A/600V fast blow ceramic fuse on mAinput
10Ainput:10Afor60 secondsmaximum followedby a
10 minutescooling period.
RESISTANCE
Ranges: 200Ω, 2k, 20k, 200k, 20MΩ
Resolution: 0.1Ω
Accuracy:
±(1.5% rdg + 4 dgts) on 200Ωto 200kΩranges
±(3.0% rdg + 5 dgts) on 20MΩrange
Open circuit volts: 0.3Vdc typical, (3.0Vdc on 200Ωrange)
Overload protection: 500VDC orAC rms
CONTINUITY
Audible indication: Less than 100Ω
Response time: 100ms
Overload protection: 500VDC orAC rms
DIODE TEST
Test current: Approx. 1.0mA
Accuracy: ±(3.0% rdg + 3dgts)
Open circuit volts: 3.0Vdc typical
Overload protection: 500VDC orAC rms
BATTERY TEST
Ranges: 1.5V, 9V
Resolution: 1mV, 10mV
Accuracy: ±(3.5% rdg + 2 dgts)
Loaded current:150mAtypical for1.5V range, 5mA
typical for 9V range
Overload protection: 500V DC orAC rms on 9V range,
0.5A/500V fast blowceramic fuseon 1.5V range.
OPERATION
Before taking any measurements, read the Safety Information Section.Always
examine the instrument for damage, contamination (excessive dirt, grease, etc.)
and defects. Examine the test leads for cracked or frayed insulation. If any
abnormal conditions exist do notattempt to make anymeasurements.
Voltage Measurements
1.Connect the red test lead to VΩjack and the black test lead to the COM
jack.
2.Set the Function/Range switch to the desired voltage type (AC or DC) and
range. If magnitude of voltage is not known, set switch to the highest range
and reduce until a satisfactory reading is obtained.
3.Connect the test leads to the device or circuit being measured.
4. For dc, a (-) sign is displayed for negative polarity; positive polarity is implied.
Current Measurements
1.Connect the red test lead to the (uA, mA or 10A) jack and the black test lead to
the "COM" jack.
2.Set the Function/Range switch to the DC orAC ranges.
3.Remove power from the circuit under test and open the normal circuit path
where the measurement is to be taken. Connect themeter in series with the
circuit.
4.Apply power and read the value from the display.
Resistance and Continuity Measurements
1.Set the Function/Range switch to the desired resistance range or continuity
position.
2.Remove power from the equipment under test.
3.Connect the red test lead to the “VΩ jack and the black test lead to
the COM jack.
4.Touch the probes to the test points. In ohms, the value indicated in the display
is themeasured value of resistance. In continuity test, the beeper sounds
continuously, if the resistance is less than 100Ω.
Diode Tests
1.Connect the red test lead to the “VΩ jack and the black test lead to
the COM jack.
2.Set the Function/Range switch to the position.
3.Turn off power to the circuit under test. External voltage across the
components causes invalidreadings.
4.Touch probes to the diode.A forward-voltage drop is about 0.6V (typical for a
silicon diode).
5.Reverse probes. If the diode is good, “OLis displayed. If the diode is shorted,
000” or another number is displayed.
6. If the diode is open, OL” is displayed in both directions.
Battery Test
1.Connect the red testleadto the (mA/BATT1.5V) jack and theblack test lead to
the "COM" jack.
2.Set the Function/Range switch to the desired 1.5V battery test range.
3.Connect the test leads to the 1.5Vdc battery under test. Normally, a good
1.5Vdc battery will read above 1.25Vdc. Consult the battery manufacturer for
complete battery specifications to determine actual battery life remaining and
condition of battery.
MAINTENANCE
WARNING
Remove test leads before changing battery or fuse or performing
any servicing.
Battery Replacement
Power is supplied bya 9 volt battery. (NEDA1604, IEC 6F22). The
" "
appears
on the LCD display when replacement is needed. To replace the battery, remove
the three screws from the back of the meter and lift off the front case. Remove
the battery from case bottom.
Fuse Replacement
If no current measurements are possible. check for a blown overload protection
fuse. For access to fuses, remove the three screws from the back of the meter
and lift off the front case. Replace F1 only with the original type 0.5A/500V, fast
acting ceramic fuse, 6.35x32mm.
Replace F2 only with the original type 10A/600V, fast acting ceramic fuse,
6.35x25.4mm.
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not use abrosives or
solvents. Dirt or moisture in the terminals can affect readings.
Safety: Conforms to IEC61010-1 (EN61010-1), CATII 1000V, CATIII 600V, Class
II, Pollution degree 2 Indoor use.
CATII: Is formeasurements performedon circuits directly connected tothe
low-voltage installation.
CAT III: Is formeasurements performedin thebuilding installation.
EMC: Conforms to EN61326.
The symbols used on this instrument are:
Caution, refer to accompanying documents
Equipment protected throughout by Double insulation (Class II)
Alternating current
Direct current
Ground
Seitenansicht 0
1 2 3 4 5

Inhaltsverzeichnis

Seite 1 - DIGITAL MULTIMETER

OPERATING INSTRUCTIONSMODEL 2703CDIGITAL MULTIMETERVVmAABATT 1.5VCOM V10A2703CSAFETY INFORMATIONThe following safety information must be observed to e

Seite 2 - MULTIMETRO DIGITAL

INSTRUCCIONES DE FUNCIONAMIENTOMODELO 2703CMULTIMETRO DIGITALVVmAABATT 1.5VCOMV10A2703CINFORMACION DE SEGURIDADLa siguiente información relativa a la

Seite 3 - Modèle 2703C

Manuel d’utilisationModèle 2703CMultimètre Numérique 2000 pointsVVmAABATT 1.5VCOM V10A2703CPRESCRIPTIONS DE SECURITELes prescriptions de sécurité ci d

Seite 4 - DIGITAL-MULTIMETER

BEDIENUNGSANLEITUNGDIGITAL-MULTIMETERMODELL 2703CVVmAABATT 1.5VCOMV10A2703CSICHERHEITSINFORMATIONENUm ein Maximum an persönlicher Sicherheit beim Betr

Seite 5 - SERVICE INFORMATION

Limited Three-Year Warranty B&K Precision warrants to the original purchaser that its products and the component parts thereof, will be free from

Kommentare zu diesen Handbüchern

Keine Kommentare