BK Precision 4011A Betriebsanleitung Seite 50

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 73
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 49
Appendix B: Equipment and Instrument Circuits
201038
Figure B.7 illustrates another thing you will need to watch out for when taking measurements
using a scope. Both Channel 1 and Channel 2 on the scope have ground leads, and those leads
are physically connected together at the scope itself. What would happen if you tried to measure
both resistor voltages using the set-up shown here? The grounds on Channel 1 and Channel 2 are
at the same potential and so would short the 200 resistor. This would again cause a large
ground loop current to flow through the scope’s ground leads, causing a circuit breaker to trip, a
fuse to blow if an autotransformer were also included in the circuit, and/or damage to the scope’s
ground leads. If you use both channels on the scope, you must make sure that the measurements
will be taken with both channel grounds at the same reference point. Slight differences between
probe ground leads that are supposed to be at the same potential will cause ground loop currents
to flow and perhaps cause errors in your results and damage to equipment.
Figure B.7 Illustration of invalid circuit measurements when using isolation transformer.
Finally, another common mistake is made in the use of the scope’s ground connections. When
taking scope measurements on an isolated circuit, you must use the probe’s ground clip from one
channel in order for the measurements to make sense. At first, it might seem that you would not
get any measurements at all if the ground connection on the probe were missing, as in Figure
B.8. However, there is always a small amount of stray capacitance in a circuit that will tie the
circuit under test back to the measuring device. Unfortunately the impedance of this ground tie is
large and it will give you unpredictable and erroneous measurements.
Figure B.8: Common problem with using isolation transformer.
Seitenansicht 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 72 73

Kommentare zu diesen Handbüchern

Keine Kommentare