
93
4. Set the external function generator for a square wave output from 0V to 2V at 200 Hz.
5. Connect the function generator output to the unit TRIG IN connector.
6. Set the waveform generator:
mode to triggered
F2 - TRIG
trigger source to
internal
7. CHECK-Verify that the oscilloscope displays one cycle of sine wave per trigger event.
8. Set the waveform generator:
to external trigger
F2-TRIG
F3 - EXT
9. CHECK-Verify that the oscilloscope displays one sine waves per external trigger
10. Set the waveform generator:
mode to burst.
MODE
F2 - BRST
F2 - INT
11. CHECK-Verify that the oscilloscope displays two cycles of sine wave per trigger event.
12. If you have a Model 4078 unit, connect the CH2 OUTPUT connector to the oscilloscope, push
CHAN key and repeat steps 2 through 13 for channel 2 (using CH2 control settings and CH2
TRIG IN connector).
5.3.8 Sync Output - Marker Output - Reference Output
Specification: TTL-level pulses at programmed frequency.
Procedure:
2) Connect the SYNC OUT connector to the oscilloscope vertical input.
3) Set the unit to generate sine waves at 1 KHz and CHECK for a square waveform with
levels of 0V and >4V. on the oscilloscope.
4) Connect the REF OUT connector to the oscilloscope vertical input.
5) CHECK for a 10 MHz square wave with 0V and >4V levels on the oscilloscope.
6) Connect the MARKER OUT connector to the oscilloscope vertical input.
7) Program the unit to generate an Arbitrary waveform with Marker:
- Press the ARB key
Kommentare zu diesen Handbüchern