the horizontal scale and position.
Turn the “time/div” knob to adjust the sampling rate
The following functions are forbidden at XY display
form:
Benchmark waveform and wave mathematics
Cursor
Auto (resets display format to YT)
Trigger Control
Horizontal Position Knob
Vector Display Type
Scan Display
Note: When the display type is set to Vector, the oscilloscope
connects the sample points by using digital interpolation.
Interpolation mode contains linear interpolation and sin(x)/x
interpolation.
Sinx/x interpolation is suitable for reconstructing
curved or irregular wave shapes, especially when the sample
rate is 3 to 5 times the system bandwidth
or at faster time base
settings.