SCPI Command Reference
98
of all programmed operations, the instrument will return to the top of the trigger model.
NOTE: With continuous initiation enabled (ON), you cannot use the :READ? Command.
:ABORt
Command Syntax:
:ABORt
Description:
When this command is sent, the multimeter aborts operation and returns to the top of the Trigger
Model.
:TRIGger
This subsystem is used for to configure trigger mode and trigger delay
:SOURce <name>
Command Syntax:
TRIGger:SOURce <name>
Command parameter:
<name> = IMMediate Internal trigger
BUS Trigger via USB/RS-232 or GPIB interface
MANual Press in the front panel for Trigger
EXTernal Trigger from external signal connected to EXT TRIG terminal
Query:
:SOURce? Query trigger source
Return String:
IMM, BUS, MAN, or EXT
Description:
This command is used to select the trigger source.
:DELay <n>
Command Syntax:
:TRIGger:DELay <n>
Command parameter:
<n> = 0 to 6000 Specify delay in milliseconds
Description:
This command is used to set the delay time for trigger delay. It takes into effect only if auto delay is
disabled.